D. TROUBLESHOOTING
(B2-BDS-3)
No.
D-1
l.
Hints
for
Using
This
Manual
This
manual
is
written
in
the
flowchart
format.
Follow
the
arrow
mark
to
find
out
a
defective
part.
The
numbers
and
symbols
like
\H-lj
and
jw-2j
indicate
the
numbers
of
connector
and
pin.
"Between
IH-11
and
IW-21"
means
"Connect
HOT
to
IH-lland
GND
to
lw-21
when
measuring
with
a
multimeter".
GND-+ A
reference
voltage,
OV,
in
an
electronic
circuit.
HOT
-+
A
line
to
apply
a
supply
voltage.
The
symbols
like
®
and
@
mean
"Jump
to
®
and
@
in
the
same
item".
The
symbols
®
and
@
mean
"Proceed
the
work
according
to
the
check
methods
®
and
@
given
in
the
same
item".
*
Be
sure
to
set
up
the
unit
before
turning
on
the
main
switch.
A
terminal
is
exposed
at
rn:=IJ
and
I
W-2~
Measure
at
the
exposed
part
without
removing.
The
electrical
base
for
BHS
is
available
in
B2-BDS-2
and
B2-BDS-3.
Refer
to
the
manuals
for
each
model.
The
difference
in
the
appearance
of
the
two
models
is
shown
below.
--------
B2-BDS-2
B2-BDS-3
LED
calor
Red
Green
Primary
voltage
YES
NO
selector
switch
lOOV
AB098800
AB615900
Nameplate
channel
AB099900
AB616100
200V
AB098900
AB6
1
6000
Channel
ABlOOOOO
AB616200
YES
The
selector
switch
is
provided.
NO
The
selector
switch
is
not
provided.
The
nameplates
are
compared
on
the
next
page.