A.
TROUBLESHOOTING
(B2-BDT)
No.
A-9
Is
power
supplied
normally?
YES
~
NO
@)
Is
the
output
of
the
transformer
TlOl
normal?
YES
NO
9
~
@)
Is
the
primary
side
input
of
the
transformer
TlOl
normal?
YES
<
NO
@
Is
the
AC
inlet
JlOl
normal?
YES
<
NO
Replace
the
JlOl
.
Is
power
supplied
normally?
YES
ubo
t~O
®
Is
the
main
switch
SWlOl
normal?
YES
NO
I
Replace
the
SWlOl.
I
l
Is
power
supplied
normally?
I
YES
~~>.N_o
____________
~
D
QV
Normal
if
AC
9V
is
generated
when
measured
with
a
multimeter
at
the
input
terminal
of
the
diode
bridge
DBlOl
.
I
Replace
I
the
TlOl.
I
I
Is
power
supplied
normally?j
@
YES
~
NO
c
Check
the
input
voltage
bet
terminal
of
the
primary
sid
selector
switch
SW102
(blue
purple
wires)
and
the
termi
the
main
switch
SWlOl
(whit
1.
When
the
input
voltage
(AC220V)
ween
the
e
voltage
and
nal
4
of
e
wire).
lS
AClOOV
Check
if
AClOOV
(AC220V)
is
detected
between
the
u p
rple
wire
of
the
SW102
and
the
terminal
4
(white
wire)
of
the
SWlOl.
2.
Check
if
AC115V (AC240V)
is
detected
between
the
blue
wire
of
the
SW102
and
the
terminal
4
(white
wire)
of
the
SWlOl.