C.
ADJUSTMENT
(B2-BDS-2)
No.
C-l
1.
Measuring
Instruments
To Be
Used
RMS
type
multimeter
(A
multimeter
capable
of
measuring
an
RMS
value)
*
An
ordinary
multimeter
can
not
be
used.
Lamp
house
with
a
12V/l00W
halogen
lamp
*
Use
a
lamp
house
with
small
contact
resistance
and
not
used
for
a
long
time.
2.
Locations
of
Adjusting
Trimmers
LED
one
RV2
RVl
UYPC65
3.
Adjustment
Procedure
®
(j)
@
Set
the
light
intensity
control
lever
to
the
"Min"
position,
and
turn
on
the
main
switch.
Set
the
light
intensity
control
lever
to
the
"Max"
position.
Turn
the
adjusting
trimmer
RV2
fully
clockwise.
Turn
the
adjusting
trimmer
RVl
fully
counterclockwise.
Measure
the
lamp
output
voltage
with
a
multimeter
(RMS
type),
and
turn
the
adjusting
trimmer
RVl
clockwise
so
that
11.8
±
0.
2V
is
obtained.
*
Measure
the
output
voltage
at
the
brass
part
for
connecting
the
lamp
house
as
illustrated
below.
type
multimeter
Turn
the
adjusting
trimmer
RV2
counterclockwise
to
set
the
output
voltage
to
11.7
±
0.2V.
Set
the
light
intensity
control
lever
to
the
"Min"
position,
and
verify
that
the
output
voltage
is
3V
or
lower.
*
Defective
if
the
voltage
is
not
lower
than
3V.
Repair
or
replace
the
PCB
UYPC65.
Turn
off
the
power,
and
wait
until
the
lamp
goes
out
completely.
Set
the
light
intensity
control
lever
to
the
"Min"
position,
turn
on
the
power
and
verify
that
the
lamp
is
lit.
Verify
that
only
the
left-most
LED
is
lit
when
the
light
intensity
control
lever
is
set
to
the
"Min"
position
and
all
the
LEDs
are
lit
when
the
lever
is
set
to
the
"Max"
position.