D. TROUBLESHOOTING
(B2-BDS-3)
No.
D-5
1
Is
the
input
voltage
of
the
secondary
PCB
Ul02
normal?
YES
·I
Replace
the
Ul02
NO
Is
the
input
voltage
of
the
display
PCB
Ul03
normal?
NO
Proceed
the
to
"No.
J-7
i YES
~I
Replace
the
display
PCB
Ul03.]
work
according
c
~
Measure
the
input
voltage
of
the
secondary
PCB
Ul02
with
a
multimeter.
Normal
if
the
input
voltage
is
as
4.
Power
Supply".
*
specified
in
the
following
table
when
the
light
intensity
control
lever
is
moved
from
"Min"
to
"Max".
Abnormal
if
not.
Input
voltage
Ul02lH-ll
gray
-
Ul02Jw-llwhite
Approx.
ACS
- 2
OV
CN3
ill
black
-
CN3
[I]
white
Approx.
AC4V
CN2
11]
purple
-
CN2
[I]
Approx.DC14
-
l7V
CN2
[l]
blue
-
CN2
[I]
Approx.DC14
-
l7V
*
Approximately
AC0.8V
is
detected
when
the
light
intensity
control
lever
is
set
to
the
"Min"
position
and
the
main
switch
is
turned
on.
This
is
not
defective
. When
the
light
intensity
control
lever
is
moved
toward
the
"Max"
direction
after
the
main
switch
is
turned
on
and
the
voltage
once
reaches
ACSV,
the
voltage
is
not
lowered
even
if
the
lever
is
moved
back
to
the
"min"
position.
"Approx.
ACS
-20V"
in
the
table
indicates
this
state.
@
Measure
the
input
voltage
of
the
dis-
play
PCB
Ul03
with
a
multimeter.
Normal
if
the
input
voltage
is
as
specified
in
the
following
table
when
the
light
intensity
control
lever
is
moved
from
"Min"
to
"Max".
Abnormal
if
not.
Input
voltage
CNl
[]]blue
-
CNl
[I]
purple
Approx.
DC
17
l4V
-
CNl
[]]gray
-
CNl
[Ij
purple
___
......
r-::;-"1
~
-
--
1-1
-
- - -
-
- -

Other manuals for Olympus BH2 Series