G. TROUBLESHOOTING
(B2-BDS-3)
No.
D-9
®
Is
the
AC
cord
(UYCP)
conducting?
@)
NO
YES
v------------~
,
Replace
the
AC
cord
(UYCPlS
in
lOOV
channel).
Is
power
supplied
normally?
~1~)
4
N~O------------~
Is
the
output
voltage
of
the
AC
inlet
PlOl
normal?
!
YES
NO
Replace
the
AC
inlet
PlOl.
Is
power
supplied
normally?
YES
-6
~~No
__
___,
Is
the
input
voltage
of
the
filter
PCB
Ul04
normal?
f
YES
0>------------1
..
NO
Replace
the
breaker
CBlOl.
~
Normal
if
the
terminals
illustrated
below
are
conducting.
@)
Measure
the
voltage
between
PlOl
~
and
PlOl
[ill
with
a
mul
timeter.
Normal
if
the
AC
line
voltage
is
detected.
Abnormal
if
not.
*
The
terminal
is
covered
with
a
tube.
Tear
it
or
puncture
with
a
needle
for
the
measurement.
Measure
the
voltage
between
Ul04
\B-2)
black
and
Ul04
\W-1\
white
with
a
multimeter.
Normal
if
the
AC
line
voltage
is
detected.
Abnormal
if
not.

Other manuals for Olympus BH2 Series