A.
TROUBLESHOOTING
(B2-BDT)
No.
A-2
2.
Lamp
2-l
Lamp
is
not
lit.
UVPC48
Causes
l.
Burnt
lamp
2.
Defective
transistor
QlOl
3.
Defective
UYPC48
4.
Defective
lamp
socket
5.
Defective
variable
resistor
RVlOl
6.
Defective
power
supply
part
0
Is
the
OV
LED
lit
when
the
main
switch
lS
turned
ON?
0
NO
Does
the
2V
-
7V
LEDs
change
YES
with
the
adjustment
of
the
0
Does
the
2V
-
7V
LEDs
light
intensity
control?
change
with
the
adjust-
YES
ment
of
the
light
Replace
the
intensity
control?
NO
y
UYPC48.
'
Proceed
the
work
according
to
"No.
A-8
4.
Power
Supply".
NO
YES
9
@
Is
the
variable
resistor
RVlOl
Remove
the
lamp
socket.
normal?
YES
NO
Is
the
lamp
conducting?
Replace
the
RVlOl.l
(Resistance
at
both
ends
of
the
lamp
~
om
f
YES
Is
the
lamp
lit?
I