23
BX3M-CB/CBFM
: Turns ON. : Blinks. :Turns OFF. : Varies depending on state.
Phenomena
Cause Remedy Page
Number
of beeps
CB HS MIX
None
Turns
ON
The light intensity
value of the MIX
slider for reflected
light observation
(U-MIXR) is 0%.
Or the hand
switch (BX3M-HS)
is damaged.
Press the light intensity button (
)
of the hand switch (BX3M-HS).
If the illumination does not turn
ON even though the light intensity
button is pressed, contact Olympus.
12
d) The OB indicator does not light even though selecting the objective by turning the motorized nosepiece.
None
The control box
(BX3M-CBFM or
BX3M-CB) is not
connected.
Set the main switch to
(OFF) and
connect the control box and the AC
adapter of the control box again.
Then, set the main switch to
(ON) and
confirm that the beep is heard once.
9, 27,
28
None
The hand switch
(BX3M-HS) is not
connected.
Set the main switch to
(OFF) and
connect the hand switch (BX3M-HS).
If the OB indicator does not light
even though the hand switch is
connected, contact Olympus.
9, 27
The hand switch
(BX3M-HS) is
damaged.
Contact Olympus.