EasyManua.ls Logo

Olympus BX3M-CB - Page 29

Olympus BX3M-CB
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
BX3M-CB/CBFM
: Turns ON. : Blinks. :Turns OFF. : Varies depending on state.
Phenomena
Cause Remedy Page
Number
of beeps
CB HS MIX
f)
The beep sound is hear
d once suddenly, and the system returns to the state that the power is ON.
1 time
The large supply
voltage fluctuation,
etc. occurs caused
by thunder, etc.
and the system
was reset.
This is not a failure.
If this phenomenon occurs
frequently, contact Olympus.
g) The beep sound is heard five times.
5 times
The
communication
between the
control box (BX3M-
CBFM or BX3M-
CB) and the hand
switch (BX3M-HS)
is disconnected.
The cable of
the hand switch
(BX3M-HS) is
removed while the
power is ON.
Set the main switch to
(OFF)
once and set the main switch to
(ON) again.
If the error occurs frequently, contact
Olympus.
9