U-B130
D.
REPAIR
PROCEDURE
3.
Firmly
tighten
the
screws
which
secure
the
left
sleeve.
Image
at
the end
of
adjustment
CHECKING
THE
LEFT/RIGHT
OPTICAL
AXIS
Change
the
interpupillary
distance
as
shown
in
Fig.
A.
Check
if
the
specimen
center
in
the
left
sleeve
is
seen
within
the
range
shown
in
Fig.
B.
Fig.
B
The
specimen
center
in
the
left
sleeve
should
be
seen
within
the
shaded
area.
Magnified
view
of
the
center
(The
center
posi-
tion
is
within
the
shaded
area.
The
digits
indicate
the
scale
gradua-
tions.)
If
the
specimen
center
is
not
seen
within
the
standard
area,
return
to
D-7
and
repeat
the
adjustment.
D-10