Olympus CX31 - Troubleshooting Guide; Optical System Troubleshooting

Olympus CX31
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CX31
16
6
Bulb is not mounted. Mount designated bulb. 6
Bulb is burnt out. Replace bulb.
Power cord plug is not connected. Connect power cord. 6
b) Bulb burns out almost immediately. Wrong type of bulb is used. Use correct bulb type. 6
Specimen is mounted upside down. Mount specimen correctly.
Cover glass is too thick. Use 0.17 mm thick cover glass.
Mount specimen correctly by
applying it on stage surface and
inserting below specimen holder.
Specimen is not correctly positioned
on the stage.
Image blurs as you move specimen.
4. Stage
a) Coarse adjustment knob is hard to
turn.
Tension adjustment ring is
overtightened.
Loosen it.
9
2. Coarse/Fine Focus Adjustment
Problem Cause Remedy Page
3. Observation Tube
b) Stage drifts down by itself or focus
is lost during observation.
Tighten it.
9
c) Coarse adjustment will not go all
the way up.
Simplified pre-focusing dial is
keeping the stage down.
Set stopper mechanism to
uppermost position.
9
d) Coarse adjustment will not go all
the way down.
Condenser holder is too low.
Raise condenser holder.
13
e) Objective makes contact with
specimen before focus is
obtained.
Specimen is mounted upside down. Mount specimen correctly.
Field of view of one eye does not
match that of the other.
Interpupillar distance is incorrect. Adjust interpupillary distance.
Your view is not accustomed to
microscope observation.
Upon looking into eyepieces, try
looking at overall field before
concentrating on specimen range.
You may also find it helpful to look
up and into distance for a moment
before looking back into microscope.
Incorrect diopter adjustment. Adjust diopter. 11
11
5. Objective Change
6. Electrical System
10
a) Bulb does not light.
Front lens of a high power objective
comes into contact with specimen
when it is engaged after a low-power
objective.
Tension adjustment ring is too loose.
10