48
IX81
Module
Specification
Operating Environment · Indoor use
· Altitude: Max. 2000 m
· Ambient temperature: 5° to 40°C (41° to 104°F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F),
decreasing linearly through 70% at 34°C (93°F), 60% at 37°C (99°F), to
50% relative humidity at 40°C (104°F)
· Supply voltage fluctuations: ±10%
· Pollution degree: 2 (in accordance with IEC60664)
· Installation (overvoltage) category: II (in accordance with IEC60664)
Objectives
Specifications
UIS
Series
For phase
contrast
observation
SLCPlanFl40XPh NA 0.55, WD 7.7 mm, ring slit IX-PH2, with correction collar
For DIC and
brightfield
observation
SLCPlanFl40X NA 0.55, WD 7.7 mm, with correction collar
For reflected
fluorescence
Fura 2
observation
UApo20X3/340 NA 0.75, WD 0.55 mm, with oil-proof cap
UApo40X3/340 NA 0.9, WD 0.2 mm, with oil-proof cap
UApo40XOI3/340 NA 1.35, WD 0.1 mm, with iris diaphragm and oil-proof cap
Water-
immersion
objectives
UApo20XW3/340 NA 0.7, WD 0.4 mm, with oil-proof cap
UApo40XW3/340 NA 1.15, WD 0.26 mm, with correction collar and oil-proof cap
UPlanApo10XW3 NA 0.4, WD 0.5 mm, with oil-proof cap