11
Installation using U-DICTS, U-DICTHC or U-DICTHR
Prism position switch lever
Pushed in: BFP1
Pulled out: NORMAL
Analyzer for transmitted
light* (U-ANT)
To be inserted into the dummy
slider of the U-KPA simplified
polarized light attachment
Transmitted light DIC slider
(U-DICTS, U-DICTHC or U-DICTHR)
DIC-compatible revolving nosepiece
(U-D6RE, U-D5BDRE, etc.)
* When the U-ANT is used in combination with the U-KPA, the field number becomes 22. If superwide-field
observation (field number 26.5) is required without using a vertical illumination, use one of the following
combinations:
@MX-AF’s optional analyzer unit MX-AFDIC + U-AN.
²U-OPA polarized light attachment + U-AN360P polarizer light rotary analyzer.
NOTE When an intermediate attachment is used, align the tube clamping knob on the micro-
scope frame with the clamping knob on the attachment in order to determine the orienta-
tion of the analyzer.
** Excessive tightening of the clamping knob hinders operation of the prism position switch lever, so tighten
the knob lightely.
Analyzer
(U-AN)
To be inserted into the
mount on the vertical
illuminator
Clamping knob**
Prism control knob
ax7353_inst.pmd 20.04.2005, 12:2711