6.4.2 JUMPER SETTING(S) FOR SENSOR EXCITATION - dc Powered Unit
Your dc-powered meter is capable of supplying either 1.25 to 12Vdc or
24 Vdc sensor excitation. (refer to Figure 6-3).
For 1.25 to 12Vdc excitation, install S1A and S4A jumpers, then adjust
the potentiometer (R34) for proper voltage.
For 24Vdc excitation, remove S4A jumpers and install S1B.
Figure 6-3. dc-Powered Main Board Jumper Positions S3 and S4
6
Jumper Positions
SEE
DETAIL A
PINS FOR RELAY
PINS FOR
ANALOG OUT
S3
S3
A
B
C
D
J12
J1
1
S
4
R
3
4
S
5
S
4
S
5
J1
0
1A
J2
A
1
10
A
AA
B
S1
6-5
S1 On dc-powered unit, side power board
S1A Installed For 1.25 to 12 Vdc excitation at 120 mA
.
S1B Installed For 24 Vdc excitation at 35 mA.
S3, S4, S5 On dc-powered unit, main board
S3A Installed To store data and setup parameters in nonvolatile memory.
S3B Omit See note in previous Section 6.4.1.
S3C Installed Unlocks lockout menu (L1 through L4).
S3D Installed Unlocks Front pushbuttons.
S4A Installed Along with the S1 jumper to program the excitation
output. Adjust excitation with R34 surface mount pot
from 1.25 to 12 volts, with an output current up to 120mA.
S4A Removed For 24 Vdc excitation.
(S4A located in storage position).
S5A Installed To enable the RESET front panel pushbutton.
S5A Removed To secure against unauthorized meter reset.