Omega DP41-S - Figure 8-6 Sensor Input Connections

Omega DP41-S
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-17
8
Scaling to
Display
Engineering
Units
STRAIN
0
DP41
METER
NC
+
E
+
S
S
E
P9
P3
1
2
3
1
2
3
GND
JUMPER
USER
PROVIDED
VOLTAGE
+EXCITATION
+OUTPUT
COMMON
Figure 8-6 Sensor Input Connections
C. The unit comes with all lockout bytes set in the unlocked
position(“0”). Refer to Section 9.1 for information on lockout
bytes.
If the unit has been programmed previously make sure that at least
the following lockout bytes are unlocked. If they are Not unlocked
you will not be able to scale the meter.
MENU SUBMENU
“L1CNF” L1C.7=0, L1C.8=0
“L2CNF” All sub menus must be unlocked “0”
Scroll to the menu by pressing the ‘MENU’ button. Enter a
submenu by pressing the ‘MIN’ button. select “0” or “1” by pressing
the ‘MAX’ button. Always press the ‘MENU’ button when leaving
a submenu to store the values you entered.
METHOD I: SCALING WITHOUT SIMULATING A KNOWN LOAD.
NOTE: All of the following steps are accomplished by using the
front panel buttons.
1. Scale the instrument by entering only 3 parameters.
The 3 parameters are “RDG SC” (Reading Scale), “RDG OF”
(Reading Offset), and “DEC PT” (Decimal Point).
Calculate “RDG SC” using the following formula: