8-19
8
Scaling to
Display
Engineering
Units
STRAIN
0
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past
L2CNF Skip past
L3CNF Skip past
L4CNF Skip past
INPUT BRIDGE Meter’s input.
RDG.CNF RDG.1=0 Scaling y=mx+b
RDG.2=0 Active decimal point
RDG.3=0 Normal display brightness
RDG.4=1 Leading zeros suppressed
RDG.5= Disregard not used
RDG.6= Disregard not used
RDG.7=0 External hard reset vs
peak reset
RDG SC 0.10000 See calculations above
RDG OF 0000000
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S1A jumper omitted)
INP.3=0 Unipolar input
(S1B jumper omitted)
INP.4=0 Std, for voltage inputs
INP.5= Disregard not used
INP.6=0 Disables IN.SC.OF.
INP.7=0 Non ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F Sets decimal point
CNT BY Press ‘RESET’ two times
to return to the run mode