8-25
8
Scaling to
Display
Engineering
Units
STRAIN
0
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
INP.3=0 Unipolar input
(S2B jumper omitted)
INP.4=0 Std, for current inputs
INP.5= Disregard not used
INP.6=0 Disables IN.SC.OF
INP.7=0 Non ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point
will be entered later)
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
2. Now you are in the run mode and the transducer is connected to
the meter. Simulate two loads on the transducer nd record the
meters display as “INPUT1” and “INPUT2” and record what
each of these reading (“READ 1” and “READ 2”) should display
on a piece of scrap paper or in the spaces provided below. These
numbers will be the INPUT values entered in the “IN.SC.OF”
scaling. Normally the two points that are calibrated are the no
load condition and 3/4 full scale. However, you may use any two
measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load and the display indicates
“9949.5”, and when you simulate 2.0000 inches of water and the
display indicates “36614.2”.