8-27
8
Scaling to
Display
Engineering
Units
STRAIN
0
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past
L2CNF Skip past
L3CNF Skip past
L4CNF Skip past
INPUT Skip past
RDG.CNF Skip past
RDG SC 0.00100 Enter new scale
value from above.
RDG OF Skip past
(Leave as 000000)
IN CNF INP.6=1 Activates “IN.SC.OF”
(2-point scaling)
IN.SC.OF INPUT 1 09949.5 This is where you
READ 1 00000.0 enter the data
INPUT2 336614.2 obtained during the
READ 2 02000.0 calibration run
(see above).
DEC PT FF.FFFF Correct decimal
point selection
CNT BY Press the ‘RESET’
button two times to
return to the run mode
The unit is now completely scaled. If during verifying the
calibration the display is not correct you can use the offset function
(“RDG OF” if all values read high or low) or perform this calibra-
tion again.