8-31
8
Scaling to
Display
Engineering
Units
STRAIN
0
After making all your choices, in submenu 1, enter these selections
by pressing the ‘MENU’ button which will momentarily flash
“STORED” on the display.
If the item selected in submenu 1 has a submenu 2 associated with
it, press the ‘MIN’ button to go to submenu 2. Select your choice
by pressing the ‘MAX’ button and enter your choice by pressing
the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one
‘MENU’ item, pressing the ‘RESET’ button two times returns you
to the run mode (this also results in a hard reset of peak and valley
and your setpoints).
Proceed through the flow chart below entering the values listed in
the Sub Menu 1 and 2 columns.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past
L2CNF Skip past
L3CNF Skip past
L4CNF Skip past
INPUT BRIDGE Meter’s input.
RDG.CNF RDG.1=0 Scaling y=mx+b
RDG.2=0 Active decimal point
RDG.3=0 Normal display brightness
RDG.4=1 Leading zeros suppressed
RDG.5= Disregard not used
RDG.6= Disregard not used
RDG.7=0 External hard reset vs
only peak reset
RDG SC 0.06250 See calculations above
RDG OF 0000000
continued next page