8-48
STRAIN
0
8
Scaling to
Display
Engineering
Units
Proceed back into the set up mode and follow the flow chart below
entering the values in the submenus shown.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past
L2CNF Skip past
L3CNF Skip past
L4CNF Skip past
INPUT Skip past
RDG.CNF Skip past
RDG SC Skip past
(Leave as 1.0000)
RDG OF Skip past
(Leave as 000000)
IN CNF INP.6=1 Activates IN.SC.OF
(2-point scaling)
IN.SC.OF INPUT 1 000013 This is where you
READ 1 000000 enter the data
INPUT2 080567 obtained during the
READ 2 000160 calibration
run (see above)
DEC PT FFFFF.F Decimal point
selection
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
The unit is now completely scaled. If during verifying the
calibration the display is not correct you can use the offset function
(“RDG OF” if all values read high or low) or perform this calibra-
tion again.