EasyManua.ls Logo

Omega DPi8 - Reading Adjust Menu; Figure 3.7 Flow Chart for Reading Adjust

Omega DPi8
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
3.2.8 Reading Adjust Menu
Figure 3.7 Flow Chart for Reading Adjust Menu
ENTER READING ADJUST MENU:
Press
a
1) Press
a
, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
INPT
Input Menu.
Press
a
3) Press
a
, if necessary, until Display advances to
R.ADJ
Reading Adjust Menu.
READING ADJUST VALUE SUBMENU:
Press
d
4) Display flashes 1
st
digit of previous Reading Adjust value.
Press
b
&
c
5) Press
b
and
c
buttons to enter a new Reading Adjust value
(-1999 to 9999).
Press
d
6) Display shows
STRD
stored message momentarily and then
advances to
SP.DV
Setpoint Deviation Menu.
Reading Offset Adjust allows the user to fine tune a minor error of the
transducer, however some applications may require a large offset adjust.
(Displayed Process Value = Measured Process Value
±
R.ADJ).
Reading Adjust value is adjustable between -1999 to 9999.
3.2.9 Setpoint Deviation Menu / Field Calibration
Figure 3.8 Flow Chart for Setpoint Deviation Menu / Field Calibration
ENTER SETPOINT DEVIATION MENU:
Press
a
1) Press
a
, if necessary, until
CNFG
prompt appears.
Press
d
2) Display advances to
INPT
Input Menu.
Press
a
3) Press
a
, if necessary, until Display advances to
SP.DV
Setpoint Deviation Menu.
For
Temperature
Reading only,
not Process

Table of Contents

Related product manuals