EasyManuals Logo

Omicron C-Probe 1 User Manual

Default Icon
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
AC/DC Current Clamp C-Probe 1
10
3.2 General Data
Dimensions: 230 x 67 x 36 mm
Max. wire diameter: 11.8 mm
Max. opening of clamp jaws: 15 mm
Connecting cable: 2 m length, isolated banana plugs
Weight: 330 g (battery included)
3.3 Input and Output Quantities
Maximum system voltage of live wire: V
eff
= 600 V with reference to ground
Bandwith of frequency: DC up to 10 kHz.
Table 3-1:
Accuracy data
Switch position 100 mV/A 10 mV/A
Measurement
range
1
1
A higher modulation leads to the saturation of the core and hence to faulty measurements.
This does not affect safety.
0-10 A
AC/DC
0-40 A
AC/DC
40-60 A
AC/DC
60-80 A
AC/DC
Accuracy
10-100Hz
1% of
meas. value
±5 mA
2% of
meas. value
±30 mA
0/-8% of
meas. value
±30 mA
0/-12% of
meas. value
± 30 mA
Accuracy
DC
1,5% of
meas. value
±5 mA
2% of
meas. value
±30 mA
0/-10% of
meas. value
±30 mA
0/-10% of
meas. value
± 30 mA
Accuracy
10 Hz - 1 kHz
1,5% of
meas. value
±5 mA
2% of
meas. value
±30 mA
0/-10% of
meas. value
±30 mA
0/-12% of
meas. value
± 30 mA
Accuracy
10 Hz - 10 kHz
1.5/-7% of
meas. value
±5 mA
0/-8% of
meas. value
±30 mA
0/-10% of
meas. value
±30 mA
0/-12% of
meas. value
± 30 mA
Phase shift
45 - 55Hz
< 0.5° < 0.5° < 0.5° < 0.5°
Phase shift
10 Hz - 1 kHz
< 2° < 2° < 3° < 3°
Phase shift
10 Hz - 10 kHz
< 30° < 30° < 30° < 30°

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron C-Probe 1 and is the answer not in the manual?

Omicron C-Probe 1 Specifications

General IconGeneral
BrandOmicron
ModelC-Probe 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals