AC/DC Current Clamp C-Probe 1
12
3.7 Environmental Conditions
Figure 3-2:
Environmental conditions
for C-Probe 1
3.8 Safety
3.8.1 Electrical Safety
Device with double isolation in hand-held area at normal usage, and
single isolation between primary winding and secondary output.
High voltage test (according to IEC 1010-1)
- 5550 V (50/60 Hz) between hand-held area at normal usage and
primary winding / secondary output.
- 3250 V (50/60 Hz) between primary winding and secondary output.
Maximum voltage (according to IEC 1010-1)
- 600 V in installation category III, degree of pollution 2.
- 300 V in installation category IV, degree of pollution 2.
Environm ental conditions
0
10
20
30
40
50
60
70
80
90
100
-50 0 50 100
Tem perature / °C
Relative humidity / %
Environmental conditions
0
20
40
60
80
100
-50 0 50 100
Temperature / °C
Relative humidity / %
Reference area
Operation
Storage

Questions and Answers: