Technical Data
3.8.2 Electromagnetic Compatibility
Electrostatic discharge: 4 kV without interference.
8 kV without destruction.
Interference field: 3 V/m according to IEC 801-2.
Transients: 1 kV class 1 without interference.
2 kV class 2 with low interference
according to IEC 801-4.
Electrical shock: 1 kV - 0,5 kA class 2 without
interference.
2 kV - 1 kA class 3 with low
interference according to IEC 801-5.
3.8.3 Mechanical Safety
Saftey class: IP 20 according to IEC 529.
Shock resistance: up to 100 g, half-period of 6 ms
according to IEC 08-2-27.
Free fall: from an altitude of 1 m according to
IEC 1010-1.
Vibration resistance: 10/55/10 Hz, 0,15 mm stroke
according to IEC 68-2-6.
Vibration test: 40 g, 6 ms, 40.000 strokes according
to IEC 68-2-29.
Clamp jaws and housing: inflammability grade V0; flame
retarding according to UL 94.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron C-Probe 1 and is the answer not in the manual?

Omicron C-Probe 1 Specifications

General IconGeneral
BrandOmicron
ModelC-Probe 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals