Technical Data
The above mentioned measurement deviations relate to the following data:
• ambient temperature of 18 °C to 28 °C
• relative humidity between 20 % and 75 %
• wire centred and parallel to the marking on the clamp
• sinus signal form for AC measurement
• impedance of the connected measuring device >100 k Ohm
• battery voltage 8 V
DC
± 0.5 V
• the operational condition of the clamp one minute after it was switched
on.
3.4 Typical Noise Levels at Output (RMS)
Table 3-2:
Typical noise levels
3.5 Influencing Factors
• Position of wire in the clamp: Max. ± 0.5 % of displayed value.
Wire from stop to stop.
• Temperature variations:
± 800 ppm/°C max. within range of
-10°C up to +50°C.
3.6 Voltage Supply
• Battery: 9 V Alkali block battery, type 6 LR 61
• Battery operating period: 70 h typically, at least 40 h
• Current consumption: 6 mA typically, 12 mA max.
• Control indicator: Green LED, off at a battery voltage
<6,5V.
Frequency band DC up to 1 Hz 1 Hz up to 10 kHz
Range 10 mV/A
15
μV
140
μV
Range 100 mV/A 1.5 mV 14 mV