AC-DC Current Clamp C-Probe 1
26
Figure 7-2:
Maximum phase error at
measurement range
100 mV/A for
measurement currents of
0-10 A (AC)
Figure 7-3:
Maximum amplitude error
at measurement range
10 mV/A for measurement
currents of 0-40 A (AC/DC)
The given accuracy data refer to the measured value. An additional
measurement error of ± 30 mA must be taken into account.
Maximaler Phasenfehler im Meßbereich 100 m
/
A(0-10A)
-35
-30
-25
-20
-15
-10
-5
0
5
10 100 1000 10000
Frequency
/
Hz
Max. error / degree
+0°/-2° +0°/-0.5° +0°/-2° +0°/-30°
Maximum phase error at measurement range 100 mV/A (0-10 A)
Maximaler
A
mplitudenfehler im Meßbereich 10 m
V
/
A(0-40A)
-10
-8
-6
-4
-2
0
2
4
1 10 100 1000 10000
Frequency
/
Hz
Max. error / %
DC
+2%/-2%
+0%/-8%
+2%/-2%
Maximum amplitude error at measurement range 10 mV/A (0-40 A)