Changing Your Pod 3
42
3 Changing Your Pod
9. Remove the needle from the vial.
Fill, Pair, and Apply the Pod
Warning: Before lling a Pod, ensure that no other Pods are being activated
within 5 feet (1.5 m) of your PDM.
Caution: Be sure to insert the ll syringe into the ll port and not any other
location on the Pod. Do not insert the ll syringe more than once into the ll port.
Only use the ll syringe and needle that came with your Pod. e ll syringe is
intended for single use only and should only be used with the Omnipod DASH™
System.
Fill the Pod with insulin
To ll the Pod with insulin (screen step ):
1. Locate the arrow on the underside of the Pod. e arrow points to the insulin
ll port.
Tip: You can leave the Pod in its tray
during lling and pairing.
2. Insert the ll syringe straight down—
not at an angle—into the ll port.
3. Depress the ll syringe plunger to
transfer the insulin into the Pod.
4. Listen for two beeps from the Pod
during the lling process. Be sure to
completely empty the ll syringe, even
aer hearing the two beeps.
Note: e Pod must contain a minimum
of 85 units of insulin in order to function.
e Pod beeps twice aer it has been lled
with 85 units of insulin. If you have lled
the Pod with more than 85 units and still
do not hear the two beeps, please call
Customer Care.
Note: Aer lling the Pod, continue to the next step immediately. If two hours
pass before pairing the lled Pod to the PDM, the Pod becomes unusable.