ERROR INDICATORS
27
Release your hand from the air
inflation bulb before the pressure
reaches 299 mmHg.
Refer to page 17.
The arm cuff was
inflated above 299
mmHg.
Press air release button and
restart the measurement with a
higher inflation level.
Refer to page 17.
Cuff is under inflated.
Replace the arm cuff with a new
one.
Air is leaking from
the arm cuff.
Remove any clothing interfering
with the arm cuff.
Clothing is interfering
with the arm cuff.
SYMBOL CAUSE CORRECTION
Movement during
measurement.
Repeat measurement. Remain
still and do not talk during
measurement. Refer to page 19.
Air tube
disconnected.
Insert the air tube securely.
Refer to page 15.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to page 15.
STORAGE
Keep the unit in its storage case when not in use.
1. Unplug the air tube from the air connector.
2. Gently fold the air tube into the arm
cuff.
NOTE: Do not bend the air tube
excessively.
3. Place the arm cuff, inflation bulb and
the main unit in the storage case.
Do not store the monitor in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
• Locations exposed to chemicals or corrosive vapours.