31
SPECIFICATIONS
Name: OMRON Manual Inflation Blood Pressure Monitor
Model: M3
Display: LCD Digital Display
Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40-180/min.
Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Inflation: Manual by inflation bulb
Deflation: Automatic pressure release valve
Rapid Air Release: Manual by push button valve
Pressure Detection: Capacitive pressure sensor
Measurement Method: Oscillometric method
Pulse Wave Detection: Capacitance type pressure sensor
Power Source: 4 “AAA” batteries
Battery Life: Approx. 1500 uses when used twice a day
with new “AAA” alkaline batteries
Operating Temperature 10ºC to 40ºC /30 to 85% RH
/Humidity:
Storage Temperature -20ºC to 60ºC /10 to 95% RH
/Humidity:
Main Unit Weight: Approximately 210 g not including batteries
Main Unit Dimensions: Approx. 106 mm (w) × 80 mm (h) × 136 mm (l)
Cuff Size: Approximately 146 mm (w) × 446 mm (l)
Cuff tube 600 mm (l)
Cuff Circumference: Fits arm circumferences 220mm to 320mm
Contents: Main Unit, Arm Cuff, Storage Case,
Four “AAA” Alkaline (LR03) Batteries,
Instruction Manual
NOTE: These specifications are subject to change without notice.
Nothing happens
when you press the
buttons.
TROUBLESHOOTING TIPS
30
Replace all four “AAA”
batteries with new ones.
Refer to page 9.
The batteries are
empty.
The unit loses power
during measurement
or the date and time
are reset.
The batteries are
worn or have been
removed for a long
time.
Replace all four “AAA”
batteries with new ones.
Refer to page 9.
The batteries have
been inserted with
their polarities
incorrectly aligned.
Check the battery
installation for proper
placement of the battery
polarities.
Press the ON/OFF button and repeat
measurement.
If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your OMRON service representative.
Other problems.
REMEDYPROBLEM CAUSES
HEM-4011C-C1 05.8.30 10:56 AM Page 30

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron M3 and is the answer not in the manual?

Omron M3 Specifications

General IconGeneral
ModelM3
Measurement MethodOscillometric
Memory Capacity60 readings
DisplayLCD
Irregular Heartbeat DetectionYes
Hypertension IndicatorYes
TechnologyIntelliSense Technology
Clinically ValidatedYes
Measurement Range (Pressure)0-299 mmHg
Cuff Size22–42 cm
Power Source4 x AA batteries
Average ReadingYes
Measurement Range (Pulse)40-180 beats per minute
Weight (without batteries)280 g
Dimensions107 mm x 79 mm x 141 mm