Detailed information about your system
8
System Accuracy Results across the
entire Glucose Range
Percent (and number) of meter results that match the
laboratory test
Within ±0.67 mmol/L or ±15%
98.2% (589/600)
Therefore, 98.2% of the total results obtained with the OneTouch®
Verio™IQ System meets this tighter accuracy specification and
exceeds the goal suggested by diabetes experts.
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch® Verio™IQ
System compares well with a laboratory method.
# of
Subjects
# of
Tests
Slope Intercept
(mmol/L)
100 600 0.948 0.263
95% CI
Slope
95% CI
Intercept
(mmol/L)
Std. Error
(S
y.x
)
(mmol/L)
R
2
0.941 to
0.955
0.181 to
0.344
0.546 0.991