92
Troubleshooting
7
What it means
One of the following may
apply:
Not enough blood or control
solution was applied or more
was added after the meter
began to count down.
The test strip may have been
damaged or moved during
testing.
The sample was improperly
applied.
There may be a problem with
the meter.
What to do
Repeat the test with a new test
strip; see blood application
(pages 36–39) or control
solution testing (pages 4552).
If the error message appears
again, contact the OneTouch®
Customer Care Line at
1 800 663-5521.
Error 4
Strip fill problem.
Retest with a
new strip.