104
Detailed information about your system
8
Precision
Within Run Precision (300 Venous Blood Tests)
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coefficient
of Variation
(%)
2.2 2.35 0.07 2.86
5.6 5.41 0.10 1.89
7.2 7.23 0.13 1.91
11.1 10.89 0.22 2.02
19.4 18.34 0.41 2.24
Results show that the greatest variability observed between
test strips when tested with blood is 2.86% or less
Total Precision (600 Control Solution Tests)
Glucose
Level
Ranges
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coefficient
of
Variation
(%)
Low (2.11–3.44) 2.79 0.10 3.52
Normal (5.67–7.67) 6.50 0.15 2.31
High (16.56–22.39) 19.00 0.48 2.53