Capture setting in image
It is possible to write brightness and diopter settings to image.
Instant review
Instant review can be ON/OFF according to user need, as available in FA Optics menu. When the reviewing
zoom level is 1x, user can change the image to be reviewed with left and right arrow keys. Images are in
showing order Normal, Low-red, Red-free, IR.
NOTE: Camera stores selected menu settings when it is powered off.
Menu:
1. Default Display window when FA optics is attached using Manual Mode. You can adjust the
diopter scale according to patient diopter manually. Image can be taken when shutter button is
pressed all the way down.
2. op:FA → Optics recognized as FA for fundus angiography imaging.
3. Default Display window when FA optics is attached using AF Assist mode. Diopter scale is
automatically adjusted, Image can be captured when shutter button is pressed half a way.
4. Default Display window when FA optics is attached using MF Assist mode. Diopter scale is
manually adjusted, Image can be captured when shutter button is pressed half a way.
5. When Right Soft key is long pressed, Menu table appears on display.
Default Device tab will be ON. By using left and right arrow keys you slide to the next tab.
Time Counter:
Start study can be selected from the FA optics menu or by pressing shutter button halfway (half-press is
enabled) or by pressing shutter button all the way down. Time counter will appear on the upper right on
the screen.
Time counter will be printed on the final images. Time counter will guide you through all the phases of
Angiography.
To stop the time counter after finishing the study, user need to go to FA optics menu and press the STOP
STUDY button.