Appendix A - Electromagnetic compatibility information
Special precautionary measures apply to this device with regard to electromagnetic compatibility (EMC).
If the performance of Optomed Smartscope M5 is lost or degraded due to electromagnetic disturbances,
it may cause unexpected or adverse operation of this device. If a malfunction cannot be rectified, please
contact your local distributor or Optomed customer service.
Use of this equipment adjacent to or stacked with other equipment should be avoided because
it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally. The device should not
be operated in the vicinity of high-frequency surgical equipment.
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the Smartscope
M5, including cables specified by the manufacturer. Otherwise, degradation of the performance
of this equipment could result.
Manufacturer’s declaration – electromagnetic immunity:
Smartscope M5 maintains basic safety and performance when used in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such an environment.
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2 kV, ±4 kV, ±6 kV,
±8 kV indirect
contact
±2 kV, ±4 kV, ±6 kV,
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for AC power
supply
Mains power quality should be that of a
typical commercial or hospital
environment.
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV differential
mode
±2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0 % U
T
for 0,5 at 8 ф
angles;
0 % 1 cycle;
70 % U
T
for 25 cycles;
0 % for 5 sec
0 % U
T
for 0,5 at 8 ф
angles;
0 % 1 cycle;
70 % U
T
for 25 cycles;
0 % for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
Smartscope M5 requires continued
operation during power mains
interruptions, it is recommended that the
device be powered from an
uninterruptible power supply or a battery.