6. Usage environment requirements
WARNING
Smartscope is not suitable for use in the presence of flammable anesthetics.
WARNING
Place PC and cradle outside the patient environment (at least 1.5 meters distance from
the patient).
CAUTION
Smartscope is intended to be used inside in a normal room temperature and normal
humidity. Do not use Smartscope in an environment where there is possibility that
water condenses to or inside the Smartscope.
CAUTION
If the PC is connected to a customer network make sure to apply appropriate safety
measures, such as Antivirus and Firewall protection.
CAUTION
It is only allowed to attach USB cable and Power source provided in the sales package
into the cradle. If you need replacement to the USB cable or Power source please
contact Manufacturer or your own Retailer.
USB Cable must be connected only to the USB port of a PC computer that complies
with the IEC 60950 standard. Avoid using excessive power or twisting the connector
when connecting the USB cord to a PC.
To transfer patient image data the device must be used together with a PC (Personal Computer) running
either Microsoft Windows 7®, Windows 8.1® and Windows 10®. The device does not need any drivers to
be installed on a PC.
Writing to the memory card from PC is not enabled by default.
Device shall be used according to this manual, quick reference guide and / or information found on
Optomed website www.optomed.com.
Electromagnetic compatibility information and recommended separation distances between portable
and mobile RF communications equipment and the Optomed Smartscope are given in the Appendix A.

Related product manuals