ENVIRONMENTAL CONDITIONS FOR USE, STORAGE AND TRANSPORTATION
IP Code: IPX0 (Equipment not protected against the ingress of water)
Intended to be use indoors:
Temperature, use: + 10 C to 35 C
Relative humidity, use: 10 % to 80 %
Atmospheric pressure: 800 hPa to 1060 hPa
Please note also EMC information given in Annex A.
Temperature, storage - 10 C to 40 C
Relative humidity, storage: 10 % to 95 %
Atmospheric pressure: 500 hPa to 1060 hPa
NOTE: If stored over 1 month, it is recommended to remove battery. Appendix B gives instructions for
removing the battery.
Transported in protective aluminum carrying case:
Temperature: - 40 C to + 70 C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 500 hPa to 1060 hPa
Sinusoidal vibration: 10 Hz to 500 Hz: 0,5 g
Shock: 30 g, duration 6 ms
Bump: 10 g, duration 6 ms
WARNING
Do not leave the eye cup in direct sunlight as it may warm up and burn patients face
when taking image.
SERIAL NUMBERING
Year of manufacture can be found from the serial number (digits 3 to 4).
Serial number of the cradle is in a sticker attached to the underside of the cradle.
Serial numbers of the attachable optics modules are attached to the modules.
SW version is displayed in the device menu.