DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual 932502G / 0618
tecture of the DSPEC 50. Setting an integral preset to 0 counts disables the preset. The
integral preset may be set to from 0 (disabled) to 4294967295 counts. See also CLEAR_
PRESETS and SHOW_INTEGRAL_PRESET.
Non-USB Execution Errors:
%131135083<CR> The command was attempted while spectrum acquisition was
in progress. No action was taken.
SET_LENGTH_SAMPLE length
Sets the number of points in the Insight mode display to the specified value. Normally this
setting would remain at 1000.
SET_LIVE ticks
Sets the live-time counter to the specified number of ticks. The number represents live time
in units of 20 ms (50 ticks/s). Normally this value is set by the DSPEC 50 during data
acquisition. See also CLEAR_COUNTER and SHOW_LIVE.
Non-USB Execution Errors:
%131135083<CR> The command was attempted while spectrum acquisition was
in progress. No action was taken.
SET_LIVE_PRESET ticks
Sets the live-time preset to the specified number of ticks (20 ms/tick). During data acquisi-
tion when the live-time counter reaches the preset number of ticks, the preset is complete and
the acquisition is stopped. Setting a live-time preset to 0 ticks disables the preset. See also
CLEAR_PRESETS and SHOW_LIVE_PRESET.
Non-USB Execution Errors:
%131135083<CR> The command was attempted while spectrum acquisition was
in progress. No action was taken.
SET_LLD chan
Sets the lower level discriminator to chan, which must be between 0 and 16383. See also
SHOW_LLD.
SET_MDA_COEF a,b,c
Sets the coefficients in the MDA preset calculation to the specified values. A, b, and c are
floating-point values. The MDA preset checks for the following condition to be met:
94