Probe, tips, accessories
15.3.3 Use of tips
All Otodynamics probe tips are disposable and MUST be discarded after
each test. The probe coupler tubes should be visually examined for signs
of contamination and the outer parts cleaned with an antiseptic wipe. Take
care not to squeeze any cleaning uid into the tubes.
The TEOAE tip design leaves a ~0.5mm gap between the end of the
coupler tubes and the end of the tip. Therefore, the tubes should never
come into contact with the patient.
OAEs should NOT be conducted if there is evidence of uid of any kind
in the ear canal. Not only does this pose a contamination risk, but OAEs
cannot be recorded through uid.
In the event of an accident with body uids, the tip, coupler tubes and
probe body must be changed.