OAE test procedure
7.3.4 Test results
The following table lists all possible test results with the associated result
graphic and gives an explanation of the circumstances under which each
result would be shown.
Test Result Description
DPOAE Pass The data collected has met the criteria set. The
optimum test setting will depend on your application,
e.g. screening or clinical measurement
(see
11.3 Test settings)
Note: The following results will only occur if a DPOAE Pass
is not obtained
No Valid OAE The data collected has not met the set pass criteria
and the test conditions were acceptable
Noisy The noisy data collected is three times greater than
the low noise data collected
Poor Probe Fit Probe t movement is detected that results in less
than 85% of stimuli reaching calibration levels
Too Few Bands Insucient bands meet their pass criteria
Stopped Too The test has been ended manually before the
Soon required minimum amount of data has been collected