Patient Safety: Patient Safety when used with the specified power supply, Delta Electronics, Inc.,
type MDS-090AAS24:
Complies with IEC 60601-1 (3rd), Class 1, Type B; AAMI ES60601-1; CSA C22.2
NO. 60601-1-08-CAN/CSA.
EMC: IEC 60601-1-2
Patient Safety when used with the specified power supply, XP Power, type
PCM80PS24:
Complies with IEC 60601-1 (2nd), Class 1, Type B; UL 60601-1; CAN/CSA-C22.2
NO 601.1-90.
EMC: IEC 60601-1-2
16.2 ACP
Outputs
Monitor headphone
3.5 mm jack (32 .)
Boom microphone 3.5 mm jack
Desktop microphone 3.5 mm jack
Inputs
Monitor headphone
3.5 mm jack (32 .)
Boom microphone 3.5 mm jack
Desktop microphone 3.5 mm jack
Built-in monitor speaker
3.5 mm jack (8 .)
Operator interface
76 buttons (61 with built-in LEDs)
2 rotary knobs (32 steps in each rotation)
USB port connector
Type: USB device port
Compliant: USB 2.0
Speed: Full-speed (12 Mb/s)
Transport and storage
Temperature: -30°C to +60°C (-22°F to 140°F)
Air humidity: 10% to 90%, non-condensing
28 Otometrics - MADSEN Astera²