EasyManuals Logo

Owon XDM2041 Series User Manual

Owon XDM2041 Series
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
3.Functions and Operations
Data Record Function
Data record function includes manual record and auto record. You can use any or both
functions to record the data. Manual and automatic records share a table of data stored in
internal storage.The maximum number of recorded points is 1000.After collecting the data, it
can be exported to the computer.
Manual record: Press the front panel key to save current reading to the DB data.
Auto record: After setting number of readings, sample interval, press the Start softkey to
the DB data.
Manual Record
1. Collect data: Press the panel key once to save the current reading to the data
table by serial number. The icon will show up on the top of the display.
Note: The measurement function can be switched during manual record. Save only
the main display readings.
2. View the manual record: Press the front panel key, press the Manual record
softkey to display the data table. Press keys to turn the page. (When the data
table is shown, you can still save current reading by pressing the key.)
Note:
When the recording data exceeds the current range, the data will be marked as
"overload".
When the relative value operation is enabled, the recorded data is still the reading
when the relative value is closed.
3. Clear the manual record: Press the Clear soft key to clear all data in the data
table.(Note: because manual and automatic records share the same data table,
automatic readings are also erased.).
Auto Record
1. Configure the parameters: Press the front panel key, press the Auto softkey.
Press the Point softkey to specify the total number of readings to record. The range is
29

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Owon XDM2041 Series and is the answer not in the manual?

Owon XDM2041 Series Specifications

General IconGeneral
TypeDigital Multimeter
DisplayLCD
True RMSYes
Continuity TestYes
Diode TestYes
Data LoggingYes
InterfaceUSB
Frequency Range0-10MHz
Power SupplyAC Adapter or Battery

Related product manuals