6.Technical Specifications
6. Technical Specifications
Function Range Resolution Accuracy: ± (% of reading + LSB)
DC Voltage
50.000 mV 0.001 mV 0.1% + 10
500.00 mV 0.01 mV 0.025% + 5
5.0000 V 0.0001 V 0.025% + 5
50.000 V 0.001 V 0.03% + 5
500.00 V 0.01 V 0.1% + 5
1000.0 V 0.1 V 0.1% + 5
True RMS
AC Voltage
500 mV – 750 V
20 Hz – 45 Hz 1% + 30
45 Hz – 65 Hz 0.5% + 30
65 Hz – 1 kHz 0.7% + 30
DC Current
500 uA 0.01 uA 0.15% + 20
5000 uA 0.1 uA 0.15% + 10
50 mA 0.001 mA 0.15% + 20
500 mA 0.01 mA 0.15% + 10
5 A 0.0001 A 0.5% + 10
10 A 0.001 A 0.5% + 10
True RMS
AC Current
500 uA – 500 mA
/
0.5% + 20
5 A – 10 A 1.5% + 20
Resistance
500 Ω 0.01 Ω 0.1% + 10
5 kΩ 0.0001 kΩ 0.1% + 5
50 kΩ 0.001 kΩ 0.1% + 5
500 kΩ 0.01 kΩ 0.1% + 5
5 MΩ 0.0001 MΩ 0.25% + 5
50 MΩ 0.001 MΩ 1% + 10
Four-wire
Resistance
500 Ω 0.01 Ω 0.1% + 10
5 kΩ 0.0001 kΩ 0.1% + 5
50 kΩ 0.001 kΩ 0.1% + 5
Diode
3.0000 V 0.0001 V /
Continuity
1000 Ω 0.1 Ω /
Frequency
10.000 Hz – 60 MHz / ± (0.2% + 8)
Capacitance
50 nF – 500 uF
/
2.5% + 5
5 mF – 50 mF 5% + 8
Temperature
K type, PT100
Display
55,000
Record Interval
15 mS – 9999.999 S
Record Length
1,000 points