Palmdoc II - Troubleshooting Control Solution Tests

Palmdoc II
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
Owner’s Manual
Questionable Results:
If the test result falls outside the range printed
on the test strip vial, check the “Trouble-Shooting
Guide” located in the “System Troubleshooting”
section of this manual and repeat the test.
Results that fall outside of the expected control
ranges may indicate:
• An error in performing the test
• Control solution bottle not shaken well
•Expired or contaminated control solution
•Control solution that is too warm or too cold
•Test strip deterioration
•Meter malfunction
•Test strips expired
Palmdoc 2 User Man.indd 33 24/08/2017 10:49