Owner’s Manual
Caution: To avoid contaminating the control
solution with the content of the test strip,
DO NOT DIRECTLY APPLY CONTROL SOLUTION
ONTO THE TEST STRIP. See Figures 15 and 16.
STEP 5 : Check if the Test Result is in Range.
After ve (5) seconds, the control solution
test result appears on the display. Compare the
test result with the range
printed on the test strip vial.
Each vial of PalmDoc Blood
Glucose Test Strips may have
a different control solution
range. The result should fall
within the printed range on
the test strip vial.
See Figure 17.
Figure 15 Figure 16
Figure 17
Normal 5-10 mmol/L
High 11.1-16.7 mmol/L