PBI-220047-F 01/2008 COPYRIGHT © PBI-Dansensor A/S
3. Connection MAP Check 9000 - User Manual
10
3.2. MAP Check PFC
•Pump control signal for external pump control (10-32VDC)
•Alarm relay outputs
•System error / Ready relay output
•Current/ voltage outputs (option)
•CAN bus (option)
Fan/ Dust filter
Mains voltage
Mains fuse
RS232C port for serial communication with PC, etc.
ST1
ST2
Gas in
Gas out
Regulated gas in
Regulated gas out
ST1 1.2: Pump control signal, 10-32VDC input (for external pump control)
3.4: O
2
alarm 1 relay. Max. 48V, 1A
5.6: O
2
alarm 2 relay. Max. 48V, 1A
7.8: System error / Ready relay. Max. 48V, 1A
ST2: 1.2: * Reserved for current output (option) (1+, 2-)
3.4: * Reserved for CAN bus (option) (3+, 4-)
5.6: * Reserved for 24VDC outlet (option) (5+, 6-)
* Other connection is marked with text on the back of MAP Check 9000
1
MEASURED
GAS:
REGULATED
GAS:
IN
MAX 6.5 BAR
(From supply)
OUT
(To application)
10