4. Operation MAP Check 9000 - User Manual
4.9. Setting alarm relays, status relay and pump control
the function of the two alarm relays. Digit No. 3 determines the function of the status relay
and the last digit determines the point from which the pump can be controlled.
0
1
2
: Key. Pump controlled by key only
: Key/Ext. Pump controlled by key and external input
: Ext. Pump controlled by external input only
0
1
2
3
Settings 4, 5, 6 and 7 correspond to 0, 1 ,2 and 3 but the alarm is not displayed
4
5
6
7
: High O concentration alarm, inverted (Open relay switch = Alarm)
: Low O concentration alarm, inverted (Open relay switch = Alarm)
: High O concentration alarm (Closed relay switch = Alarm)
: Low O concentration alarm (Closed relay switch = Alarm)
: High O concentration alarm, inverted (Open relay switch = Alarm)
: Low O concentration alarm, inverted (Open relay switch = Alarm)
: High O concentration alarm (Closed relay switch = Alarm)
: Low O concentration alarm (Closed relay switch = Alarm)
2
2
2
2
2
2
2
2
0
1
2
3
: System error, inverted (Open relay switch = System error)
: Ready (Closed relay switch = Ready)
: System error (Closed relay switch = System error)
: Ready, inverted (Open relay switch = Ready)
Alarm 1 Alarm 2
Pump
PA1:
Status
relay
Pump controlled by
MAP Check 9000 status relay
MAP Check 9000 alarm relay 1 & 2