PBI-220047-F 01/2008 COPYRIGHT © PBI-Dansensor A/S
4. Operation MAP Check 9000 - User Manual
16
4.9. Setting alarm relays, status relay and pump control
The parameter PA.1 is defined in the setup menu. The first two digits (from the left) determine
the function of the two alarm relays. Digit No. 3 determines the function of the status relay
and the last digit determines the point from which the pump can be controlled.
0
1
2
: Key. Pump controlled by key only
: Key/Ext. Pump controlled by key and external input
: Ext. Pump controlled by external input only
0
1
2
3
Settings 4, 5, 6 and 7 correspond to 0, 1 ,2 and 3 but the alarm is not displayed
4
5
6
7
: High O concentration alarm, inverted (Open relay switch = Alarm)
: Low O concentration alarm, inverted (Open relay switch = Alarm)
: High O concentration alarm (Closed relay switch = Alarm)
: Low O concentration alarm (Closed relay switch = Alarm)
: High O concentration alarm, inverted (Open relay switch = Alarm)
: Low O concentration alarm, inverted (Open relay switch = Alarm)
: High O concentration alarm (Closed relay switch = Alarm)
: Low O concentration alarm (Closed relay switch = Alarm)
2
2
2
2
2
2
2
2
0
1
2
3
: System error, inverted (Open relay switch = System error)
: Ready (Closed relay switch = Ready)
: System error (Closed relay switch = System error)
: Ready, inverted (Open relay switch = Ready)
Alarm 1 Alarm 2
Pump
PA1:
Status
relay
Pump controlled by
MAP Check 9000 status relay
MAP Check 9000 alarm relay 1 & 2