PBI-Dansensor A/S COPYRIGHT © PBI-220047-F 01/2008
15
MAP Check 9000 - User Manual 4. Operation
The setup menu SET.
1. Press when SET. is shown in the display.
2. Use
and to select the required value.
3. Press
for three seconds to change the value (please refer to “Setting alarm limits” on
page 13 for entry of values).
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
•FLO.L 100 Flow alarm limit. Alarm limit for min. flow in ml/min.
•
FLO.H 350 Flow alarm limit. Alarm limit for max. flow in ml/min.
•
DEL. 3 Gas flush time in seconds before measurement.
Valid range: 0 to 250.
•
PA.1 0001 Setting alarm relays 1 and 2, device status relay and pump.
See"Setting alarm relays, status relay and pump control" on page 16.
•
PA.2 1000 Setting output format and three-way valve.
See "Setting O2 output" on page 17.
•
PA.3 0000 Setting up various extra equipment.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
When entering FLO.l and FLO.H, er. 1 is presented if an attempt is made to enter values which
make
FLO.l higher than FLO.H.
The diagnostics menu DIA.
1. Press when DIA. is shown in the display.
2. Select the required value by means of
and . Press for approx. three seconds for
continuous display of the selected value. The value will be shown until
is pressed again.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
•EMF. 165.0 Current sensor EMF in mV.
•
TPT.1 33 Current internal MAP Check 9000 temperature in °C.
•
TPT.2 0 Current heater temperature difference in °C.
Difference between current temperature and ideal temperature.
•
OFFS -0.12 Sensor offset. The result of the last offset adjustment in mV.
•
ti.po 1234 Hour counter (10 hour increments). Number of hours for which the
MAP Check 9000 has been switched on.
•
ti.ca 678 Hour counter (10 hour increments). Number of hours since the last
calibration of the MAP Check 9000.
•
TI.PU 983 Hour counter (10 hour increments). Number of hours for which the
pump has been on.
•
S.ref 7.65 This menu point will only be indicated if the option "Measurement of
the O
2
sensor reference" has been installed. The output value is
stating the reference of the O
2
sensor.
See "Measurement of the Sensor Reference Value parameter of the
Oxygen Sensor" on page 21.
•
PRG. 5.3 Installed program version.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Calibration menu CAL.
See the section concerning calibration.