EasyManua.ls Logo

PBI Dansensor MAP Check 9000 - P.P.C Menu

Default Icon
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PBI-Dansensor A/S COPYRIGHT © PBI-220047-F 01/2008
27
MAP Check 9000 - User Manual 7. Options
7.2.1 . P. F. C / P. P. C menu
fl.in : Flush in flow.
Entered as
lo/Hi.xx. When online measurement is started, the system starts with flush in
flow. This flow is maintained for
del. seconds.
hi.o2/lo.o2 : Regulation window, upper/lower O
2
limit for the required O
2
value.
Together these two O
2
values form a window (upper and lower O
2
limit) within which MAP
Check tries to keep the measured O
2
value. If the O
2
value measured is higher than hi.o2,
the proportional valve is opened until the O
2
value measured is below hi.o2.
If the O
2
value measured is lower than lo.o2, the proportional valve is closed until the O
2
value measured is above or equal to
lo.o2. As long as the O
2
value measured is within the
two limit values the flow is not adjusted.
The
lo.o2 value entered cannot exceed the hi.o2 value.
reg.h/reg.l : Max./min regulation values
Entered as
lo/Hi.xx values.
These two values allow you to fix absolute limits for upward and downward regulation of
the flow.
g.up/g.dn : Flow regulation gain
The speed at which the flow rate is increased/decreased when the O
2
value measured is
outside the defined regulation window.
g.up and g.dn are entered as O
2
values in the range from 0.00ppm to 99.99% O
2
, where
1ppm corresponds to a gain factor of 1. These two values have nothing to do with O
2
, but in
order to increase the value range they are entered as O
2
values, as this allows the gain
value to be entered as a factor in the range from 0.01 to 999900.0 (0.01ppm - 99.99 %)
The reason why there is an upward factor as well as a downward factor is that a fast flow
increase is typically required when the oxygen level is too high at the measuring point.
G.D : The degree of differentiation
The degree with which the flow is increased/decreased when rapid changes appear in the
measured O
2
value. G.D is entered as O
2
values in the range from 0.00 ppm to 99.99% O
2
.
If
G.D is set to 0 the differentiation has no effect.

Table of Contents

Related product manuals