10. Specifications MAP Check 9000 - User Manual
38
10. Specifications
10.1. Sensor
Solid-state ionic conductor with stabilized zirconium oxide electrolyte.
Temperature: 1000K (727°C)
Warm-up time: 20 min., full specification after 40 min.
Calibration interval: 12 months
Reference gas: Atmospheric air (20.946 % oxygen)
Sensor response time: Less than 50 msec.
System response time: Less than 10 sec., depending on flow and measuring gas
hose
Calibration: Atmospheric air 20.946 % + up to four optional calibration
gases
10.2. Flow system, MAP Check 900X-1 and MAP Check 900X-2
The measuring gas is supplied continuously by means of an electric pump. The flow system
in MAP Check 900X-1 and MAP Check 900X-2 can avaible with or without bypass flow. A
bypass flow will give a faster response time but will consume more gas.
Fig. 5. Flow diagram, MAP Check 900X-1 and MAP Check 900X-2
O
sensor
2
Flow
sensor
By-pass
hose
Approx. 200 ml/min
0.2 - 1.2
l/min
Pump
Gas In
Gas Out