PBI-220047-F 01/2008 COPYRIGHT © PBI-Dansensor A/S
10. Specifications MAP Check 9000 - User Manual
42
10.6. Technical data
•Mains connection: 230 VAC (115 VAC) +/- 10 %, 50-60 Hz
•Consumption: Max. 160 VA, nom. 80 VA.
•Mains fuse: 1 AT at 230 V (3.15 AT at 115 V)
•Permissible ambient temperature during operation:
0 to 45°C. Less than 95%RH up to 35°C.
Linear RH reduction by 3%RH/°C between 35°C and
45°C.
•Permissible ambient temperature while stored:
-20°C to +45°C. Less than 75%RH
•Cabinet (Stand-alone): Painted ALU. cover with handles.
•Built-in cabinet: ALU. cover with mounting holes and handles.
•Dimensions: See next page
•Weight: MAP Check 9000: Approx. 8.5 kg
Transport box: Approx. 5 kg
•Installation dimensions: Panel hole: 179x236 (HxW mm)
The height corresponds to 4U
Depth: at least 477 mm
The device must be supported.
Vent holes must not be covered.
NOTE! The unit must be supported - do not cover air holes.