Prima 451 – User Manual 13
Description
3.9.3 AV-S MRI
Ventilator
Power
S
upp
l
y
The ventilator (1) power supply unit is connected internally to
the machine power supply.
Note that the Prima 451 gas delivery switch (2) must be in the
‘On’ position for operation of the ventilator (see section 5.4),
and refer also to section 3.4 in the AV-S MRI user manual.
AV-S
Ventilator
Back-up
B
a
tt
e
r
y
If the power supply to the ventilator fails, the ventilator
back-up battery will power the ventilator for 60 minutes, if the
battery has been maintained in a fully charged condition.
Refer to section 3.4 in the AV-S MRI user instruction manual.
3.9.4
Monitor
and other de
v
i
c
e
s
The mains lead (or adaptor) for a monitor system or other
devices requiring an electrical supply can be plugged into
one of the auxiliary sockets on the rear of the machine (see
section 3.9.2).
Warning
The Prima 451 MRI has been designed for use in an MRI
environment only as a system.
It is the user’s responsibility to ensure that fitting third-
party components to the anaesthesia machine, ventilator, or
breathing system will not compromise the MR status of the
Prima 451 MRI System.
Do not place any object on or in this machine unless it is
specifically labelled to be used in an MRI scanning room
and suitable for your facility.
Objects placed on or in this machine that are not designed
for use with this anaesthesia system may be strongly
attracted to the magnet and may cause serious injury or
death when the machine is used in an MRI scanning room.
If a third-party accessory or device is mounted on the
machine (e.g. AGSS receiver or patient monitoring
equipment) ensure that it is MR compliant for your facility,
and is securely mounted.
Installation must be carried out by trained engineering
personnel.
In addition, do not assume that individual components of the
Prima 451 MRI System (including any Penlon accessories),
can be safely used with MRI scanners when used on any other
anaesthesia system.
See also Warnings 1 and 2 in section 1).
3.10
Auxiliary
Gas
O
u
t
l
e
t
s
CAUTION
When the auxiliary gas outlets are in use on a machine with
cylinder supply only, or if the pipeline supply is not in use,
check flow rate requirements, and ensure that adequate
back-up cylinders are available.
Warning
Use only non-magnetic cylinders with this machine. Steel
cylinders can cause serious injury or death if brought into
an MRI scanning room.
Oxygen and
A
i
r
Auxiliary outlets (1) are mounted on the rear of the machine.
Supply
p
r
e
ss
u
r
e: See section 4.5.
1
2
1